2 Nov 2001 W. A. Traub SHORT DELAY PRESSURE SETTINGS ----------------------------- I measured optimum pressure settings for SD-1/2 in early July. The units are pounds per square inch (psi), which can be read directly on the gauges near the control knobs. The measurements were made with air in the tank, as follows: Location Pressure range (air) Pressure range (vac) SD-1 top 63 - 100 (best = 70) est. best = 70 SD-1 side 55 - 100 (best = 60) est. best = 60 SD-2 top 45 - 46 (best = 45) est. best = 45, maybe 36 SD-2 side 42 - 46 (best = 43) est. best = 43, maybe 34 The minimum pressure was found by noting at what point the force needed to pull the cars started to increase, as pressure was reduced. The minimum force, at pressures abobe the minimum, leveled out at about 0.1-0.4 kg for SD-1, and 0.1-0.2 kg for SD-2. The measurement was made with a spring scale, over a range of about 50 cm, near home, and the worst case force was noted. It may in fact be worse farther down the track, but I could not measure there. The carts should *not* be run at low pressures which might allow dragging of the air pads to occur, because this will cause deep grooves to be cut in the granite surface, and will increase the oscillation dramatically. We know this from experience. The way to be sure that we are not rubbing on the granite is, ideally, to measure the current needed to drive the cart, and to be sure that the current is never greater than some value, which we still need to determine empirically. If someone can set up these tests, I would be glad to discuss this with them; we have an ADC card which sould be used to monitor this current, or perhaps there is a better way. Ultimately it should be logged as a housekeeping parameter. The maximum pressure is that at which the cart starts to vibrate, at about 60 Hz. At pressures above this point the vibration increases linearly. The vibration was measured by looking at the oscilloscope display of the error signal, so it is a measure of the OPD noise, which is the aspect that we care most about. The noise was measured when slewing the SD at 1 cm/s. Note that an error beyond the range of +/-5 volts (I believe) on the oscilloscope is bad, because at this point the error rolls over and we pick up a permanent path error of several (10 ?) microns (MGL please fill in here). The error signal can be digitized with the ADC card and logged. Both this and the drive current should appear in a little graph that the operator can see on the VxW monitor, and a flag set if either value goes out of bounds. If vibration is a problem at all pressures, possibly especially on SD-2, then the magnetic pull-down force should be adjusted to minimize vibration. This is done by turning one of the 3 jack screws on the top (usually) of the SD platform. The dihedral will have to be removed to do this (4 screws from base plate to cart top). About 1/4 turn, either CW or CCW, should help. Try to mark where you start, so you don't get lost. A full turn is probably far too much. The reason for this problem is that the original adjustment decays, due to a poor locking mechanism on the jack screw; if this worked better, we might not lose the adjustment. In vacuum, the atmospheric pressure of about 670 torr or 9 psi is removed, so the proper flotation pressure should be less than in air by this amount. However I have not tested this in the present set-up, so the safe way to proceed would be to start in vacuum with the same pressure as in air, and reduce the pressure cautiously watching both the drag and the error measurements.