ANORAD
Last update: February 1998
-
Turn on Air to table (toggle valve on green tank, below the delay line
tank, next to the Fluor table).
-
Check air flow rate to values specified on card - about 5 liters/minute
with air filled tank, 4 liters/minute with vacuum. Higher flow rates will
induce oscillations of the carriage location and
orientation. Valves need to be adjusted when changing between vacuum and
air filled tanks.
-
Have Rate cards powered up and a delay line control program operating.
-
Command the delay lines to be idle (stopped).
-
Turn on Anorad Servo (upper blue box on the big rack).
-
Clear laser faults by pushing reset (front black button on the big rack)
on HP box.
Anorad system should now be locked up and ready to operate.